We propose and validate a semi-automatic method for lumen segmentation of the carotid bifurcation in computed tomography angiography (CTA). First, the central vessel axis is obtained using path tracking between three user-defined points. Second, starting from this path, the segmentation is automatically obtained using a level set. The cost and speed functions for path tracking and segmentation make use of intensity and homogeneity slice-based image features. The method is validated on a large data set of 234 carotid bifurcations of 129 ischemic stroke patients with atherosclerotic disease. The results are compared to manually obtained lumen segmentations. Parameter optimization is carried out on a subset of 30 representative carotid bifurcations. With the optimized parameter settings the method successfully tracked the central vessel paths in 201 of the remaining 204 bifurcations (99\%) which were not part of the training set. Comparison with manually drawn segmentations shows that the average overlap between the method and observers is similar (for the inter-observer set the results were 92\% vs. 87\% and for the intra-observer set 94\% vs. 94\%). Therefore the method has potential to replace the manual procedure of lumen segmentation of the atherosclerotic bifurcation in CTA.
Robust CTA lumen segmentation of the atherosclerotic carotid artery bifurcation in a large patient population
R. Manniesing, M. Schaap, S. Rozie, R. Hameeteman, D. Vukadinovic, A. van der Lugt and W. Niessen
Medical Image Analysis 2010;14(6):759-769.